10K Closed Cycle Cryostat

CCS (Closed Cycle System) refrigerator utilizes the principle of G-M thermodynamic cycle to compress and expand helium. The gas is expanded and cooled in the cold head, and the expanded low-pressure helium gas is returned to the compressor, and the cycle is repeated. The refrigerator consists of two stages, the first stage is used to cool the heat shield around the sample, and the second stage is used to cool the sample.
  •  

            CCS (Closed Cycle System) refrigerator utilizes the principle of G-M thermodynamic cycle to compress and expand helium. The gas is expanded and cooled in the cold head, and the expanded low-pressure helium gas is returned to the compressor, and the cycle is repeated. The refrigerator consists of two stages, the first stage is used to cool the heat shield around the sample, and the second stage is used to cool the sample.

            The 10K closed cycle cryostat provided by Jinzhengmao Technology Co., Ltd. is a variable temperature and low temperature platform based on G-M refrigerators produced by domestic and foreign companies. Warm, easy to operate, and low maintenance cost.

    Scope of use:It is mainly used for magnetic measurement, optical measurement, photoelectric reaction, light reflection, fluorescence and magnetoresistance in low temperature and vacuum environment.

    Sdvantage:

    ※ Compact structure, high cost performance, no refrigerant consumption when reaching low temperature, long maintenance period, strong expansibility, and easy to use;

    ※ The sample is in a high vacuum environment or atmosphere;

    ※ Can be integrated with optical windows, a variety of optional window materials;

    ※ A variety of gold-plated oxygen-free copper sample holders can meet the measurement needs of different experimental requirements, and the sample holders can be quickly replaced;

    ※Suitable for electrical, optical and magnetic measurements in low temperature environments;

    ※Silicon diode thermometer or Cernox thermometer (used in strong magnetic field) is installed on the cold finger;

    ※The compressor can be air-cooled or water-cooled, and the maintenance cycle is long;

    ※A variety of options can meet the measurement needs of different experiments.

    The standard configuration:

    ※ A set of vacuum hood

    ※ One temperature sensor and one temperature control heater

    ※ A gold-plated oxygen-free copper sample holder

    ※ Vacuum port KF25

    ※ Air release valve KF16

    ※ An 8pin electrical vacuum connector, a temperature sensor and a heater share a 6-pin connector

    ※ Matching compressor, cold head drive line, high pressure helium gas pipe and Toolkit.

    ※ A set of standard cold head placement bracket

    Use:

    ※ It is mainly used for magnetic measurement, optical measurement, photoelectric reaction, light reflection, fluorescence and magnetoresistance in low temperature and vacuum environment, including material research, superconductivity, cold pump, cold screen and helium liquefaction test.

    ※ Matrix isolation

    ※ High temperature (700K) DLTS test

    ※ Mössbauer spectrum

    ※ High Tc

    ※ PL spectrum

    ※ Photoelectric effect

    ※ UV/VIS/IR spectrum test

    Optional:

    ※ Rotary vacuum hood

    ※ Atmosphere introduction end

    ※ Larger sample space

    ※ Various low temperature electrical leads (coaxial wire, tee shaft, phosphor bronze twisted pair, microwave semi-steel cable, etc.)

    ※ DLTS four-probe sample holder

    ※ Electrical connectors (BNC, SMA, SMK and multi-pin connectors)

    ※ Special shaped window (such as concave window), fifth window (bottom window)

     

    Sumitomo 10K Closed Cycle Cryostat

    Commonly Used 10K Refrigerator Cold Head And Compressor Technical Specifications And Parameters

    Cold Head Model

    CH-202

    CH-204

    CH-204NOptional

    KDE210SA

    Low Temperature As Low As

    10K

    10K

    6.5K

    10K

    Cooldown

    90min to 20K

    40min to 20K

    60min to 10K

    60min to 10K

    Cooling Capacity

    1.8W@20K&7.3K@77K

    6.7W@20K&13.5W@80K

    2.5W@20K&

    2.5W@20K&40W@45K

    Cold Head Weight

    6.8Kg

    7.8Kg

    7.8Kg

    17.8Kg

    Compressor

    HC-4Ewater cooling 

    KDC6000Vwater cooling

    Power Consumption

    2.6KW at 50Hz/1Phase

    7.2KW at 50Hz/3Phase

    Cooling Water Inlet Flow And Temperature

    2.7L/min@4-27℃

    7-10L/min@28℃

    Size

    504*430*485mm

    613*442*600

    Compressor Weight

    75Kg

    118Kg

    Maintenance Cycle

    30 000h

    36 months

     

    Integration Specifications And Parameters

        Model

    Project

    CH-202

    CH-204

    CH-204NOptional

    KDE210SA

    Temperature Range

    10K-325K

    Required Vacuum

    ≤1Pa

    Experiment Type

    Electricity

    Optics

    Electricity

    Optics

    Electricity

    Optics

    Sample Holder Size

    15*25mm

    15*25mm center opening Φ10mm light hole

    20*30mm

    20*30mm center open Φ10mm light hole

    15*25mm

    15*25mm center opening Φ10mm light hole

    Vacuum Hood Size

    Φ31.8mm

    40*40mm

    Φ41.8mm

    50*50mm

    Φ31.8mm

    40*40mm

    Sample Holder Terminal

    8

    None

    8

    None

    8

    None

    Temperature Sensor/Heater

    The standard configuration of the temperature sensor is a silicon diode, and the standard configuration of the heater is 50W

    Optical Window

    None

    1-5

    None

    1-5

    None

    1-5

    Note: Optical holes can be opened in the middle of the optical sample holder, and the size can be specified by the user.

    Sample Holder Options:

    Model

    Specifications And Parameters

    Illustrate

    Remark

    T8015

    15*25mm

    Additional electrical sample holder with 8 binding posts above and below.

    1. 1. All standard 10K thermostats are equipped with a sample holder by default. The size can be made according to the standard, or it can be designed according to the specific requirements of the user. If the design is more complicated, it can be charged as appropriate;
    2. 2. The standard optical sample holder comes with a sample tablet with a hole on the tablet, the diameter of which is the same as that of the clear hole;
    3. The cooling power of the 10K refrigerator is slightly different, and the size of the sample holder can be adjusted within a certain range according to the specific power.

    T8020

    20*30mm

    Additional electrical sample holder with 8 binding posts above and below.

    T8030

    30*35mm

    Additional electrical sample holder with 8 binding posts left and right.

    T8040

    40*45mm

    Additional electrical sample holder with 8 binding posts left and right.

    T8115

    15*25mm

    Additional optical sample holder with Φ8 through hole in the middle.

    T8120

    20*30mm

    Additional optical sample holder with Φ10 through hole in the middle.

    T8130

    30*35mm

    Additional optical sample holder with Φ15 through hole in the middle.

    T8140

    40*45mm

    Additional optical sample holder with Φ20 through hole in the middle.

    T8205

    Φ5mm

    Powder sample container with 5mm window on one side of the container.

    T8210

    Φ10mm

    Powder sample container with 5mm window on one side of the container.

     

     

    Optical window options:

    Window Model

    Window Specification

    Illustrate

    Remark

    W-20

    Φ20*1.5mm

    Additional Quartz Windows

    1. 1. The specifications of the window are all visible area, and the actual size is larger than the visible area;
    2. For window installation, you can choose O-ring compression or low-temperature epoxy resin bonding. If the size of the vacuum cover is small, try to use glue.

    W-40

    Φ40*2mm

    Additional Quartz Windows

    K9-W20

    Φ20*1.5mm

    Additional K9 window

    K9-W20

    Φ40*2mm

    Additional K9 window

    F-W20

    Φ20*1.5mm

    Additional calcium fluoride window

    F-W40

    Φ40*2mm

    Additional calcium fluoride window

    Cable Options:

    Name

    Illustrate

    Remark

    PTFE Wire

    Ordinary wire, large heat leakage, diameter between 0.1-0.3

    The cable is not required to use a common PTFE wire alone, it is required to be made according to the requirements, and the wire should be used in conjunction with the corresponding connector

    Low Temperature Coaxial Cable

    Low signal interference and accurate measurement

    Phosphor Bronze Twisted Pair

    Poor thermal conductivity, small heat leakage, diameter 0.2mm (4 pieces)

    Microwave Semi-rigid Cable

    High frequency cable, high frequency signal use

     

    Connector Options:

    Electrical Connector Model

    Specifications And Parameters

    Describe

    T91-8P Ordinary Glass Sintering

    8pin

    Add a quick-plug 8P common electrical connector.

    T91-12P Ordinary Glass Sintering

    12pin

    Add a quick-plug 12P common electrical connector.

    T91-SMA

    0-18GHz

    With Flexible Cable

    0-12.4GHz

    With semi-rigid, semi-flexible cable

    T91-BNC

    0-4GHz

    With Ordinary Coaxial Cable

    T91-TRX

    Triax Connector

    Add a quick-plug triax connector.

    T91-S

    Stainless Steel Junction Box

    Add a stainless steel junction box.

    T91-PTEF

    PTFE Wiring Cover

    Add a Teflon junction cover to the stainless steel junction box.

     

    Remark:
    1: All thermostats are shipped with two connectors, one is for temperature sensor and heater installation, and the other is for other performance measurements. When more test wires are required, it is necessary to increase Mounting holes, or add one or more Breakout Boxes.
    2: There is T91-S option in the above table, the terminal board and the shell of this option constitute the grounding as a whole, and the other is the T91-PTEF option, this option is the shell and the installed joint using PTFE board insulation connection, The electrical properties of the housing and the connector are independent of each other. This should be taken into account when selecting models.

    Water cooler selection

    There are split and integrated options, please refer to the water cooler color page for details.

    The standard configuration of mechanical pump and molecular pump is a mechanical pump. If higher vacuum is required, domestic and foreign molecular pumps can be selected. For selection, refer to the vacuum pump selection table.

     

     