PT100 Platinum Resistance Temperature Sensor
-
PT100 Platinum Resistance Temperature Sensor
The PT100 platinum resistance thermometer is an excellent choice for applications that require temperature measurement and control between 70K and 500K. In this temperature range, platinum resistance thermometers have good repeatability and almost constant sensitivity. Within the allowable error range (1 degree), platinum resistance thermometers can also be used in magnetic fields and radiation environments.
Platinum resistance thermometers above 70K have a common standard curve, which makes platinum resistance thermometers very interchangeable. The thin-film platinum resistors provided by Jinzhengmao have smaller size, shorter response time, and wider application occasions.
Features
※ Above 40K is less affected by magnetic field;
※ Minimal effect of ionizing radiation; v
※ Provide two-point or three-point or multi-point calibration.
Parameter
※ Temperature range: 14K ~ 873K
※ Standard curve: IEC 751
※ Recommended excitation current: 1mA
※ Recommended power dissipation under excitation: 100μW@273K
※ Sensitivity: follow the typical sensitivity curve of PT100
※ Response time: 1.5S@77K, 10S@273K
※ Radiation effect: can be used in radiation environment
※ Magnetic field influence: It is not recommended to use in a magnetic field environment below 40K
※ Repeatability: ±10 mK@77K (from 100 thermal shock tests from 300K to 77K)
Typical Characteristic Curve







Typical Temperature And Resistance Relationship:
Temperature
R(Ω)
S(Ω/K)
20K
2.2913
0.085
50K
9.3865
0.360
77k
20.380
0.423
150K
50.788
0.409
300K
110.354
0.387
600K
221.535
0.372
800K
289.789
0.360
Calibration Accuracy:
20-77K
77-305K
305 -400K
400-475K
475-500K
2P
±0.25K
±0.9K
±1.3K
±1.4K
3P
±0.1K
±1.4K
20L
±0.1K
±0.1K
20H
±0.1K
±0.1K
±0.1K
±0.1K
±0.1K
70L
±0.1K
70H
±0.1K
±0.1K
±0.1K
±0.1K
2P:77K and 273K calibration 3P:77K, 273K and 475K calibration